Guangzhou Cleanroom Construction Co., Ltd.

GCC : You All-in-one Cleanroom Construction and Solutions Partner

Manufacturer from China
Verified Supplier
5 Years
Home / Products / Lab Room Equipment /

Digital Display Vertical Pressure Steam Sterilizer Autoclave Automatic

Contact Now
Guangzhou Cleanroom Construction Co., Ltd.
Visit Website
City:guangzhou
Province/State:guangdong
Country/Region:china
Contact Person:MsSophie
Contact Now

Digital Display Vertical Pressure Steam Sterilizer Autoclave Automatic

Ask Latest Price
Video Channel
Brand Name :GCC
Model Number :HLCB
Certification :CE,ISO
Place of Origin :China
MOQ :1 Set
Price :Negotiation
Payment Terms :T/T,
Delivery Time :7-14 Workdays
Packaging Details :Plywood Box
Keyword :pressure steam sterilizer autoclave
Type :portable vertical
Style :sterilizer autoclave
Material :stainless steel
Function :sterilization
After-sales service provided :Online support
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

Digital Display Automatic Vertical Pressure Steam Sterilizer/ Autoclave

1.Sterilization Work Process Description
1.1 Preparation
1)Open the lid of the container, take out the drums and add water manually till it reach the high level. (Use distilled water)
2) Turn on the power and turn on the power break switch.
3) Standby indicator light is on. The high water level indicator lights up.
4) After piling the items, place them on a sieve plate sequentially, and leave some appropriate gaps between the packages. We suggest to make the packages into 20cm x 20cm x 10cm. The dressing and the textile should not be tightly packed. And don’t forget to place the sterilizing indicator.
5) After placing the drum in the container, close the and clockwisely tighten the hand wheel till the lid is closed. Don’t close too tight as which might bring damage to the rubber gasket.
6) Set sterilization parameters (see parameter settings for details).

The Sterilizing time can be set by the followed table 2, please set the sterilizing time and temperature according to different item required.

CHARACTERISTICS

  • Equipped with a pressure/temperature controller, the pressure setting range is 0.07 to 0.22MPa

(corresponding to a saturated steam temperature of 115 to 134°C).

  • Equipped with a timer to automatically control the sterilization time.
  • Equipped with a safety valve, the relief valve would automatically releases the pressure when the

sterilization pressure exceeds the maximum allowable value.

  • The immersed type electric heating tube, which is with anti-dry function. When the water level is

lower than the specified water level line, the sterilizer automatically cuts off the heating power and alarms at the same time.

Parameters

Digital Display Vertical Pressure Steam Sterilizer Autoclave Automatic